FDA Adverse Event
Malfunction
Summary report: N
I-STAT TROPONIN (CTNI) CARTRIDGE
MDR report key: 2210185
·
Received May 9, 2011
Report
- Report Number
- 2245578-2011-00062
- Event Type
- Malfunction
- Date Received
- May 9, 2011
- Date of Event
- April 18, 2011
- Report Date
- May 6, 2011
- Manufacturer
- ABBOTT POINT OF CARE CANADA LTD.
- Product Code
- MMI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED REGARDING AN I-STAT TROPONIN CARTRIDGE THAT YIELDED AN UNEXPECTED RESULT OF 0.14 NG/ML ON A PT WHEN RUNNING LITHIUM HEPARIN PLASMA. REPEAT TEST RESULT ON THE SAME SAMPLE RUNNING WHOLE BLOOD USING THE I-STAT WAS 0.03 NG/ML. BASED ON THE INFO AVAILABLE THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE CANADA LTD. | NA | T11040 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |