FLOGARD
Report
- Report Number
- 6000001-2011-18648
- Event Type
- Malfunction
- Date Received
- August 12, 2011
- Date of Event
- July 18, 2011
- Report Date
- July 18, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS REPAIRED ON-SITE AT THE CUSTOMER FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE A FAULTY MAIN BATTERY. TO CORRECT THE CONDITION, THE MAIN BATTERY WAS REPLACED.? IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
(B)(4). THE DEVICE WAS INVESTIGATED BY A BAXTER FIELD SERVICE ENGINEER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP REQUIRED REPLACEMENT OF A (B)(4) BATTERY. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |