ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-05835
- Event Type
- Malfunction
- Date Received
- August 12, 2011
- Date of Event
- July 16, 2011
- Report Date
- July 16, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE UNRESPONSIVE; INVESTIGATORS WERE UNABLE TO MOVE BEYOND THE VERIFY SCREEN DUE TO BUTTON UNRESPONSIVENESS. THERE WAS EVIDENCE OF MOISTURE BEHIND THE DISPLAY LENS. EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE USER GUIDE WARNS THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND/OR THE WATERPROOF FEATURE OF THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.
THE PATIENT REPORTED THAT ON (B)(6) 2011 SHE NOTICED THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE, AND THEN AFTER SWIMMING IN THE POOL THE BUTTONS WERE COMPLETELY UNRESPONSIVE. SHE STATED THAT SHE REBOOTED THE PUMP IN AN ATTEMPT TO SOLVE THE ISSUE, AND THE BUTTONS REMAINED UNRESPONSIVE. THE PATIENT DENIED PHYSICAL DAMAGE TO THE RUBBER KEYPAD, AND DENIED VISIBLE MOISTURE IN THE BATTERY COMPARTMENT OR BEHIND THE DISPLAY SCREEN. SHE STATED THAT SHE WEARS THE PUMP IN HER POCKET AND DOES NOT CLEAN THE PUMP. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |