FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2210120 · Received August 12, 2011

Report

Report Number
2531779-2011-05836
Event Type
Malfunction
Date Received
August 12, 2011
Report Date
July 16, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES AND REQUIRED INCREASED FORCE TO ENGAGE. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. UNRELATED TO THE COMPLAINT, THE BOLUS BUTTON WAS FOUND TO BE DETACHED, THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED, AND THE DISPLAY SCREEN WAS FOUND TO BE FADED AND DISCOLORED.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE KEYPAD BUTTONS HAVE BEEN DIFFICULT TO PRESS FOR THE PAST FEW MONTHS. HE CONFIRMED THAT THE KEYPAD IS INTACT. HE STATED THAT HE WEARS THE PUMP IN HIS POCKET AND DOES NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 22 YR