COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-18621
- Event Type
- Malfunction
- Date Received
- August 12, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 8, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE, BUT THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE BE RECEIVED FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). EVALUATION SUMMARY:THE CUSTOMER'S REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM WAS DETERMINED TO BE DEPLETED BATTERIES RESULTING FROM USER ERROR. THE BATTERY AND BATTERY HARNESS WERE REPLACED TO FIX THE REPORTED CONDITION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. IT IS UNKNOWN IF THERE WAS REPORTED PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. THE SOFTWARE VERSION IS CURRENTLY UNKNOWN AT THIS TIME. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |