FDA Adverse Event Malfunction Summary report: N

OLYMPIC COOL-CAP SYSTEM

MDR report key: 2209848 · Received August 2, 2011

Report

Report Number
3018859-2011-00003
Event Type
Malfunction
Date Received
August 2, 2011
Date of Event
July 7, 2011
Report Date
August 2, 2011
Manufacturer
NATUS MEDICAL INCORPORATED
Product Code
MXM
PMA / PMN Number
P040025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

COOL-CAP SYSTEM IS BEING RETURNED TO NATUS MEDICAL INC FOR TESTING. A F/U REPORT WILL BE SUBMITTED TO THE FDA WHEN THE DEVICE INVESTIGATION / TESTING IS COMPLETED.

Description of Event or Problem · 1

DURING TREATMENT OF AN INFANT, AN OLYMPIC COOL-CAP SYSTEM EXHIBITED A SHUTDOWN CHARACTERIZED BY A FROZEN DISPLAY SCREEN. NO AUDIBLE OR VISIBLE ALARMS WERE OBSERVED. SYSTEM WAS FROZEN FOR ONE HOUR AND 44 MINUTES BEFORE IT WAS DISCOVERED. RECTAL TEMPERATURE WAS 36.9 CENTIGRADE WHEN ISSUE WAS DISCOVERED. PT WAS MOVED TO A SECOND COOL-CAP DEVICE. THE REMAINDER OF THE 3-DAY TREATMENT WAS COMPLETED WITHOUT FURTHER INCIDENT. THERE WAS NO PT INJURY. DURING INITIAL ON-SITE DEVICE TESTING, THE HOSP'S BIOMEDICAL TECH OBSERVED A FEW GREEN LINES AT THE BOTTOM OF THE SCREEN AND THE DEVICE REPEATEDLY REBOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OLYMPIC COOL-CAP SYSTEM COOLING CAP (INFANTS) MXM NATUS MEDICAL INCORPORATED 60010 NOT APPLICABLE

Patients

Seq Age Sex Outcome Treatment
1