FDA Adverse Event Malfunction Summary report: N

ELECSYS TROPONIN T HS STAT

MDR report key: 22095183 · Received May 28, 2025

Report

Report Number
1823260-2025-01628
Event Type
Malfunction
Date Received
May 28, 2025
Date of Event
May 8, 2025
Report Date
July 10, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
MMI
PMA / PMN Number
K201441
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 0

THE SERIAL NUMBER OF THE E601 ANALYZER IS (B)(6). THE FIELD SERVICE ENGINEER CHECKED THE ANALYZER FLUIDICS. THE TEST WAS ASSIGNED TO THE OTHER MEASURING CELL OF THE ANALYZER. CALIBRATION AND CONTROLS WERE RUN. PATIENT SAMPLES WERE TESTED IN DUPLICATE AND THE RESULTS WERE COMPARABLE. THE INVESTIGATION IS ONGOING.

Description of Event or Problem · 0

ROCHE DIAGNOSTICS RECEIVED A CUSTOMER COMPLAINT REGARDING ALLEGED NON-REPRODUCIBLE ELECSYS TROPONIN T HS STAT ASSAY RESULTS FOR THREE PATIENT SAMPLES TESTED ON A COBAS 6000 E601 MODULE. THE CUSTOMER NOTED THEY WERE HAVING CALIBRATION ISSUES. AFTER PERFORMANCE OF MAINTENANCE, CALIBRATION AND CONTROLS WERE ACCEPTABLE. THE FIRST SAMPLE INITIALLY RESULTED IN A TROPONIN T VALUE OF 8.27 NG/L AND WHEN REPEATED ON A SECOND ANALYZER, THE RESULT WAS 22.12 NG/L. THE SECOND SAMPLE INITIALLY RESULTED IN A TROPONIN T VALUE OF 8.68 NG/L AND WHEN REPEATED ON A SECOND ANALYZER, THE RESULT WAS 14.44 NG/L. THE THIRD SAMPLE INITIALLY RESULTED IN A TROPONIN T VALUE OF 94.32 NG/L AND WHEN REPEATED ON A SECOND ANALYZER, THE RESULT WAS 241 NG/L.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
327973 ELECSYS TROPONIN T HS STAT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ROCHE DIAGNOSTICS 827211

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown