FDA Adverse Event Malfunction Summary report: N

ULTIPACE¿ LEAD, 52CM MRI

MDR report key: 22093946 · Received May 28, 2025

Report

Report Number
3008377825-2025-00004
Event Type
Malfunction
Date Received
May 28, 2025
Date of Event
May 8, 2025
Report Date
June 16, 2025
Manufacturer
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
Product Code
NVN
UDI-DI
05415067046314
PMA / PMN Number
P960013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT OF HELIX MECHANISM ISSUE WAS CONFIRMED BUT HELIX DAMAGE WAS NOT CONFIRMED. AS RECEIVED, A COMPLETE LEAD WAS RETURNED IN ONE PIECE WITH THE HELIX PARTIALLY EXTENDED AND CLOGGED WITH BLOOD/TISSUE. VISUAL AND X-RAY EXAMINATION OF THE LEAD DID NOT FIND ANY ANOMALIES. ELECTRICAL TESTING DID NOT FIND ANY INDICATION OF CONDUCTOR FRACTURES OR INTERNAL SHORTS. AFTER CLEANING THE HELIX, THE HELIX COULD BE EXTENDED AND RETRACTED BY APPLYING TORQUE TO THE CONNECTOR PIN. THE MEASURED FULL HELIX EXTENSION LENGTH WAS WITHIN SPECIFICATION. THE CAUSE OF HELIX MECHANISM ISSUE WAS ISOLATED TO THE HELIX BEING CLOGGED WITH BLOOD/TISSUE.

Additional Manufacturer Narrative · 0

CORRECTION: D3 - THE CORRECT MANUFACTURER ESTABLISHMENT NAME SHOULD BE ST. JUDE MEDICAL, INC.(CRM-SYLMAR) INSTEAD OF ST. JUDE MEDICAL OPERATIONS (M) SDN BHD. G8 - MANUFACTURER REPORT NUMBER SHOULD HAVE STARTED WITH 2017865 NOT AS SUBMITTED 3008377825-2025-00004.

Additional Manufacturer Narrative · 0

FURTHER INFORMATION WAS REQUESTED BUT NOT RECEIVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT PRESENTED FOR AN IMPLANT PROCEDURE. IT WAS NOTED THAT THE HELIX OF THE RIGHT ATRIAL LEAD WAS DAMAGED. IT WAS ALSO NOTED THAT THE HELIX FAILED TO BE EXTENDED AND RETRACTED. THE LEAD WAS NOT USED AND REPLACED DURING THE PROCEDURE. PATIENT STATUS WAS NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332158 ULTIPACE¿ LEAD, 52CM MRI NO MATCH NVN ST. JUDE MEDICAL, INC.(CRM-SYLMAR) LPA1231/52 P000226718 05415067046314

Patients

Seq Age Sex Outcome Treatment
1 91 YR Female