CD HORIZON® SPINAL SYSTEM
Report
- Report Number
- 1030489-2025-02094
- Event Type
- Malfunction
- Date Received
- May 28, 2025
- Date of Event
- May 2, 2025
- Report Date
- July 9, 2025
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- HXX
- UDI-DI
- 00643169043640
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- 003
Narratives
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
H3: PRODUCT ANALYSIS # (B)(4): PART # 5584111, LOT # K24H1080 VISUAL INSPECTION CONFIRMED THE ENTIRE TORX TIP OF INSTRUMENT HAS BEEN SHEARED OFF. OPTICAL EXAMINATION OF THE FRACTURE SURFACE REVEALED A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW. THE DAMAGE TO THE DRIVER IS CONSISTENT WITH TORSIONAL OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, USER FACILITY) REGARDING A DRIVERS HAVING SPINAL THERAP Y. IT WAS REPORTED THAT TIP OF THE VOYAGER DRIVER BROKE OFF AND THE TIP OF THE SLEEVE DRIVER BROKE OFF. PRODUCTS WERE USED NUMEROUS TIMES. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 330331 | CD HORIZON® SPINAL SYSTEM | SCREWDRIVER | HXX | MEDTRONIC SOFAMOR DANEK USA, INC | 5584111 | K24H1080 | 00643169043640 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |