FDA Adverse Event Malfunction Summary report: N

BABY CONTROL COSYCOT INFANT WARMER

MDR report key: 2208518 · Received August 11, 2011

Report

Report Number
9611451-2011-00483
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
July 5, 2011
Report Date
July 20, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
FMT
PMA / PMN Number
K971695
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT WARMER HEAD HAS NOT BEEN RETURNED TO THE MANUFACTURER. IT HAS BEEN VISUALLY INSPECTED BY A FISHER & HEALTHCARE SERVICE PERSONNEL AT OUR REGIONAL OFFICE IN (B)(4). THE VISUAL INSPECTION CONFIRMED THE REPORTED FAULT OBSERVED BY THE CUSTOMER. MULTIPLE CRACKS AND CHIPS WERE FOUND ON THE SIDES OF THE UPPER MOLDED HEATER HEAD CASE. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS TYPE FOR LOT 020924. IT IS LIKELY THAT THE CRACKING OF THE COMPLAINT WARMER HEAD IS RELATED TO A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE (B)(4) COMPONENTS OF THE INFANT WARMER. WHEN THE WARMER HEAD BEGINS TO WARM UP DURING USE, A CHEMICAL REACTION WITH THE INCOMPATIBLE CLEANING SOLUTION RESULTS IN GRADUAL CRACKING AND CRAZING OF THE WARMER HEAD. THE INFANT WARMER SERVICE MANUAL FOR THE AFFECTED UNIT FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS (B)(4) COMPONENTS AND STATES THE FOLLOWING: CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTERALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE; CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES. AS PART OF CONTINUOUS IMPROVEMENT, WE HAVE SINCE REVISED OUR CLEANING INSTRUCTIONS TO RECOMMEND SPECIFIC PROPRIETARY CLEANING WIPES. THE COMPLAINT WARMER HEAD CASING HAS SINCE BEEN REPLACED WITH A NEW UPPER HEAD CASING AND RETURNED TO THE HEALTHCARE FACILITY.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THE HEATER HEAD OF AN (B)(4) BABY CONTROL COSYCOT INFANT WARMER WAS CRACKED. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BABY CONTROL COSYCOT INFANT WARMER FMT FMT FISHER & PAYKEL HEALTHCARE LTD IW930

Patients

Seq Age Sex Outcome Treatment
1