FDA Adverse Event Malfunction Summary report: N

BD CONNECTA PLUS3 RED

MDR report key: 22084714 · Received May 26, 2025

Report

Report Number
9610847-2025-00156
Event Type
Malfunction
Date Received
May 26, 2025
Date of Event
April 13, 2025
Report Date
June 5, 2025
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMG
UDI-DI
00382903946051
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 394605 AND LOT NUMBER 4124010 . THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD CONNECTA PLUS3 RED WAS BLOCKED WITH MINOR INJURY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MILD INJURY THE PATIENT WAS PUMPED WITH PRESSOR DRUGS FOR 24 HOURS, AND THE THREE-WAY VALVE WAS BLOCKED AND DRUG-INTOLERANT. INFUSION OF LIQUID MEDICINE. IMMEDIATELY REPLACE THE NEW CONSUMABLE THREE-WAY AND CONTINUE TO INFUSE THE LIQUID.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
728458 BD CONNECTA PLUS3 RED STOPCOCK, I.V. SET FMG BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 4124010 00382903946051

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown