FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 22081626 · Received May 23, 2025

Report

Report Number
9617229-2025-08614
Event Type
Injury
Date Received
May 23, 2025
Date of Event
November 2, 2016
Report Date
May 23, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
UDI-DI
10888628000308
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS IS A FOLLOW-UP REPORT TO A MEDWATCH SUBMITTED UNDER MANUFACTURE REPORT NUMBER 9617229-2025-0000036. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION. DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: USE ERROR: OBSERVED, ONE OPENING ASSESSED AS SURGICAL DAMAGE. DEFLATION: OBSERVED, ONE OPENING ASSESSED AS FOLD FLAW OPENING. AS PER THE INVESTIGATION PROCEDURE CREASE, NON-PENETRATING NICK AND WEAR ABRASION WERE OBSERVED. NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. IT WAS ADDITIONALLY REPORTED "HOLE ON BOTTOM". THE DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
326425 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 2881900 10888628000308

Patients

Seq Age Sex Outcome Treatment
1 38 YR Female Required Intervention