TECNIS ODYSSEY TORIC II IOL
Report
- Report Number
- 3012236936-2025-000147
- Event Type
- Injury
- Date Received
- May 23, 2025
- Date of Event
- April 23, 2025
- Report Date
- January 9, 2026
- Manufacturer
- AMO MANUFACTURING NETHERLANDS
- Product Code
- MFK
- UDI-DI
- 05050474811416
- PMA / PMN Number
- P980040
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CORRECTED DATA: UPON FURTHER REVIEW IT WAS NOTED THAT SECTION G4: PMA/510(K) NUMBER IN THE INITIAL MDR WAS INADVERTENTLY SUBMITTED MISSING THE ¿P¿ BEFORE THE NUMERICAL VALUE. THE COMPLETE NUMBER IS (B)(4). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
SECTION A3B, A4 AND A5: UNKNOWN/ ASKED BUT NOT AVAILABLE. SECTION E1: SURGEON'S EMAIL ADDRESS: UNKNOWN/ ASKED BUT NOT AVAILABLE. SECTION H3: OTHER 81: THE DEVICE WAS NOT RETURNED FOR EVALUATION. THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORD AND COMPLAINT TRENDING FOR THIS DEVICE WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ATTEMPTS HAVE BEEN MADE TO OBTAIN MISSING INFORMATION; HOWEVER, TO DATE, NO RESPONSE HAS BEEN RECEIVED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
THE PRODUCT WAS RETURNED TO THE MANUFACTURING SITE FOR EVALUATION. ADDITIONAL INFORMATION: SECTION D9: DEVICE AVAILABLE FOR EVALUATION? YES. SECTION D9: DATE RETURNED TO MANUFACTURER: JUN 2, 2025. SECTION H3: EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: VISUAL INSPECTION UNDER MAGNIFICATION WAS PERFORMED ON THE SUSPECT PRODUCT AND FOUND THE LENS WAS COATED IN VISCOELASTIC RESIDUE AND FIBERS FROM THE MEDICAL GAUZE. THE LENS WAS CLEANED AND INSPECTED REVEALING DAMAGE ON THE SURFACE OF THE LENS. NO FURTHER ISSUES WERE OBSERVED. NO FURTHER EVALUATION WAS PERFORMED. NO ISSUES THAT COULD CAUSE OR CONTRIBUTE TO THE COMPLAINT ISSUES REPORTED WERE IDENTIFIED DURING PRODUCT EVALUATION. THE OTHER ISSUES OBSERVED DURING THE PRODUCT EVALUATION COULD NOT BE CONFIRMED TO BE RELATED TO THE MANUFACTURING OR DESIGN PROCESS CONCLUSION: BASED ON THE COMPLAINT INVESTIGATION RESULTS, THE PRODUCT WAS RELEASED WITHIN SPECIFICATIONS. NO PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION COULD BE IDENTIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM PATIENT'S LEFT EYE AS THE PATIENT WAS INTOLERANT OF MULTI-FOCAL SIDE-EFFECTS. THE LENS WAS EXPLANTED AND REPLACED WITH A NON JNJ LENS IN A SECONDARY PROCEDURE. THERE WAS NO MEDICAL/SURGICAL INTERVENTION REQUIRED. FROM ADDITIONAL INFORMATION IT WAS LEARNT THAT THERE WAS NO PATIENT INJURY. NO OTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 961623 | TECNIS ODYSSEY TORIC II IOL | LENS, MULTIFOCAL INTRAOCULAR | MFK | AMO MANUFACTURING NETHERLANDS | DRT150 | 05050474811416 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Female | Required Intervention |