FDA Adverse Event Malfunction Summary report: N

CERENOVUS ENTERPRISE

MDR report key: 22078066 · Received May 23, 2025

Report

Report Number
3008114965-2025-00472
Event Type
Malfunction
Date Received
May 23, 2025
Date of Event
May 19, 2025
Report Date
June 16, 2025
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
NJE
UDI-DI
10886704044024
PMA / PMN Number
H60001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER REF#: (B)(4) UPDATED SECTIONS ON THIS MEDWATCH: B4, G3, G6, H2, H3, H6 AND H11. COMPLAINT CONCLUSION: AS REPORTED BY THE FIELD, DURING AN ENDOVASCULAR EMBOLIZATION, AN ENTERPRISE® VASCULAR RECONSTRUCTION DEVICE (ENC452212, 8872358) BECAME IMPEDED IN THE UNSPECIFIED MICROCATHETER HUB AND COULD NOT ADVANCE ANY MORE. THE DOCTOR ONLY RETRACTED THE STENT ALONE, BUT THE STENT WAS FOUND DETACHED FROM THE DELIVERY WIRE IN THE PLACE BETWEEN THE MICROCATHETER HUB AND Y CONNECTOR. THE DOCTOR REMOVED THE Y CONNECTOR AND REMOVED THE STENT AND THEN SWITCHED TO A NEW STENT TO COMPLETE THE SURGERY. THE MICROCATHETER (MC) WAS NOT REPLACED. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONAL EVENT INFORMATION RECEIVED ON 22-MAY-2025 INDICATED THAT THEY WERE ABLE TO TORQUE THE DEVICE. THERE WAS NO EVIDENCE OF PHYSICAL MATERIAL WITHIN THE DEVICE. THE MICROCATHETER DID NOT KINK/BENT. THERE WAS NO EXCESSIVE FORCE USED WITH THE DEVICES. THERE WERE NO PROCEDURAL DELAY/PROLONGATION DUE TO THE EVENT. ONE PHOTO ACCOMPANIED THE COMPLAINT FILE. IN SAID PHOTO, A STENT COMPONENT CAN BE SEEN OUTSIDE THE INTRODUCER AND DETACHED FROM THE DELIVERY WIRE. THE STENT CAN BE SEEN FULLY EXPANDED, AND NO DAMAGES ARE OBSERVED. THE DEVICE WAS RETURNED TO J&J MEDTECH FOR FURTHER EVALUATION. A NON-STERILE EU 4.5X22MM STENT 12 MM DW TIP WAS RECEIVED CONTAINED IN THE DECONTAMINATION POUCH. UPON RECEIVING THE DEVICE, VISUAL INSPECTION WAS PERFORMED, AND THE STENT WAS DETACHED FROM THE DELIVERY UNIT. MICROSCOPIC INSPECTION WAS PERFORMED ON THE STENT COMPONENT. IT WAS OBSERVED TO BE IN GOOD CONDITION; THERE WAS NO STRUCTURAL DAMAGE (I.E., NO BROKEN STRUTS, NO KINKS); ALSO, IT WAS NOTED FULLY EXPANDED, BOTH ENDS CAN BE NOTED AS COMPLETELY FLARED. THE DELIVERY WIRE WAS SUBJECTED TO DIMENSIONAL ANALYSIS AND ALL MEASUREMENTS WERE FOUND TO BE WITHIN SPECIFICATION, INCLUDING THOSE SPECIFICATIONS THAT CONTROL THE ATTACHMENT AND DELIVERY OF THE STENT. THEREFORE, DEVICE FAILURE IS NOT SUSPECTED TO BE A CONTRIBUTING FACTOR. THE CUSTOMER COMPLAINT REGARDING A STENT BEING DETACHED WAS CONFIRMED SINCE THE STENT WAS NOTED AS ALREADY SEPARATED FROM THE DELIVERY SYSTEM; BASED ON THIS CONDITION, THE ISSUE REGARDING A STENT BEING IMPEDED IN THE HUB OF THE CONCOMITANT MICROCATHETER CANNOT BE EVALUATED THROUGH FUNCTIONAL TESTING. THE STENT MUST BE INSIDE THE INTRODUCER TUBE TO PERFORM THE FUNCTIONAL ANALYSIS. ADDITIONALLY, NONE OF THE RETURNED COMPONENTS PRESENT DAMAGES THAT SUGGEST THAT THEY WERE FORCIBLY ADVANCED. WITH THE LIMITED INFORMATION AVAILABLE, A CONCLUSIVE CAUSE CANNOT BE DETERMINED; HOWEVER, IT IS POSSIBLE THAT CLINICAL AND PROCEDURAL FACTORS, INCLUDING DEVICE MANIPULATION AND OPERATOR'S TECHNIQUE, MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. AT THIS TIME, THERE IS NO EVIDENCE TO SUPPORT THAT THE ISSUE REPORTED IN THE COMPLAINT IS A RESULT OF A DEFECT INHERENTLY RELATED TO THE DEVICE. LAKE REGION MEDICAL DID REVIEW THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTING AND PACKAGING OF THE LOT 8872358. THE HISTORY RECORDS INDICATE THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. AS PART OF THE J&J MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CAPA ACTIVITY IS REQUIRED. IT SHOULD BE NOTED THAT PRODUCT FAILURE COULD BE CAUSED BY MULTIPLE FACTORS. THE INSTRUCTIONS FOR USE (IFU) DO CONTAIN THE FOLLOWING RECOMMENDATIONS: DO NOT PARTIALLY DEPLOY THE STENT FROM THE INTRODUCER. IF RESISTANCE IS MET DURING MANIPULATION, DETERMINE THE CAUSE OF RESISTANCE BEFORE PROCEEDING. DO NOT APPLY UNDUE FORCE IF RESISTANCE IS ENCOUNTERED AT ANY POINT DURING STENT MANIPULATION. WITHDRAW THE UNIT AND ADVANCE A NEW ONE. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED TO IDENTIFY STATISTICAL SIGNALS FOR CONSIDERATION OF FURTHER CORRECTION ACTION. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT.

Additional Manufacturer Narrative · 0

MANUFACTURE REF#: (B)(4). INFORMATION REGARDING PATIENT WEIGHT, HEIGHT, MEDICAL HISTORY, RACE, AND ETHNICITY WAS NOT REPORTED. SECTION H3: THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION AND TESTING. HOWEVER, IT HAS NOT BEEN RECEIVED TO DATE. IF THE DEVICE RETURNS, A DEVICE INVESTIGATION WILL BE PERFORMED. MISSING INFORMATION FROM THIS REPORT IS IDENTIFIED AS BLANK; THIS INFORMATION WAS NOT PROVIDED IN THE REPORTED EVENT OR AVAILABLE AT THE TIME OF REPORT SUBMISSION. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY CERENOVUS, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, CERENOVUS, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

MANUFACTURER REF#: (B)(4). UPDATED SECTIONS ON THIS MEDWATCH: B4, D9, G3, G6, H2, H3, H6 AND H11. THE PRODUCT HAS BEEN RETURNED FOR EVALUATION AND TESTING; HOWEVER, THE ENGINEERING EVALUATION HAS NOT BEEN COMPLETED. ONE PHOTO ACCOMPANIED THE COMPLAINT FILE. IN SAID PHOTO, A STENT COMPONENT CAN BE SEEN OUTSIDE THE INTRODUCER AND DETACHED FROM THE DELIVERY WIRE. THE STENT CAN BE SEEN FULLY EXPANDED, AND NO DAMAGES ARE OBSERVED. A DEVICE HISTORY RECORD (DHR) WAS PERFORMED, AND IT INDICATED THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. THE IMPEDED CONDITION REPORTED CANNOT BE EVALUATED THROUGH A PHOTO; HOWEVER, THE DETACHMENT OF THE STENT CAN BE CONFIRMED DUE TO THE FINDINGS MENTIONED ABOVE. THIS INVESTIGATION WAS PERFORMED BASED ONLY ON THE PHOTO PROVIDED. AN ASSESSMENT WILL BE PERFORMED AS PER THE CONDITIONS OF THE DEVICE RETURNED. AS PART OF J&J MEDTECH QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CAPA ACTIVITY IS REQUIRED. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT.

Description of Event or Problem · 0

AS REPORTED BY THE FIELD, DURING AN ENDOVASCULAR EMBOLIZATION, AN ENTERPRISE® VASCULAR RECONSTRUCTION DEVICE (ENC452212, 8872358) BECAME IMPEDED IN THE UNSPECIFIED MICROCATHETER HUB AND COULD NOT ADVANCE ANY MORE. THE DOCTOR ONLY RETRACTED THE STENT ALONE, BUT THE STENT WAS FOUND DETACHED FROM THE DELIVERY WIRE IN THE PLACE BETWEEN THE MICROCATHETER HUB AND Y CONNECTOR. THE DOCTOR REMOVED THE Y CONNECTOR AND REMOVED THE STENT AND THEN SWITCHED TO A NEW STENT TO COMPLETE THE SURGERY. THE MICROCATHETER (MC) WAS NOT REPLACED. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONAL EVENT INFORMATION RECEIVED ON 22-MAY-2025 INDICATED THAT THEY WERE ABLE TO TORQUE THE DEVICE. THERE WAS NO EVIDENCE OF PHYSICAL MATERIAL WITHIN THE DEVICE. THE MICROCATHETER DID NOT KINK/BENT. THERE WAS NO EXCESSIVE FORCE USED WITH THE DEVICES. THERE WERE NO PROCEDURAL DELAY/PROLONGATION DUE TO THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
819438 CERENOVUS ENTERPRISE INTRACRANIAL NEUROVASCULAR STENT NJE MEDOS INTERNATIONAL SARL 8872358 10886704044024

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown MICROCATHETER (UNSPECIFIED BRAND).