FDA Adverse Event Malfunction Summary report: N

BD POSIFLUSH XS/SF

MDR report key: 22071748 · Received May 22, 2025

Report

Report Number
9616657-2025-00007
Event Type
Malfunction
Date Received
May 22, 2025
Date of Event
March 27, 2025
Report Date
May 22, 2025
Manufacturer
BECTON, DICKINSON AND CO.
Product Code
NGT
UDI-DI
00382903065721
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR IS THE RESULT OF AN INVESTIGATION FINDING: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 306572 AND LOT NUMBER 4242754. THE REVIEW DID NOT REVEAL ANY POSSIBLE DETECTED PRODUCT NON-CONFORMANCES DURING THE PRODUCTION PROCESS THAT COULD HAVE BEEN RELATED TO THIS REPORTED INCIDENT. TO AID IN THE INVESTIGATION OF THIS ISSUE, BOTH PICTURE AND PHYSICAL SAMPLES WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. THROUGH EXAMINATION OF THE SAMPLES, PACKAGE DAMAGE ON THE TOP WEB COMPONENT WAS OBSERVED. THIS PACKAGE DAMAGE WOULD BREACH THE STERILITY OF THE BLISTER PACK. A REVIEW OF THE MES (MANUFACTURING ENGINEERING SYSTEM) WITH THE TECHNICAL TEAM SHOWED THAT THERE HAD BEEN AN ISSUE WITH THE PACKAGING ROBOT ON A SHIFT DURING THE MANUFACTURE OF THIS BATCH WHICH MOST LIKELY CAUSED DAMAGE TO THE PACKAGING TOP WEB. THIS ISSUE WAS CONFIRMED RESOLVED ON THE NEXT SHIFT. THIS IS THE FIRST REPORT RECEIVED FOR THIS TYPE OF DEFECT ON LOT NUMBER 4242754. AT THIS TIME, FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY. OUR QUALITY TEAM WILL CONTINUE TO CLOSELY MONITOR THE MANUFACTURING PROCESS FOR SIGNS OF THIS POTENTIAL DEFECT AND ANY EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD POSIFLUSH-XS / SF HAD PACKAGE DAMAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: POSIFLUSH OUTER PACKAGE (PAPER SIDE) DAMAGE DUE THE PRESSURE COMPACTED INSIDE BOX AND POSITION OF THE SYRINGE FROM SUPPLIER. FOUND IN PPA AREA NO HARM CAUSED TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2367153 BD POSIFLUSH XS/SF SALINE VASCULAR ACCESS FLUSH NGT BECTON, DICKINSON AND CO. 4242754 00382903065721

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown