FDA Adverse Event Malfunction Summary report: N

LOCKING SCREW 40

MDR report key: 22068817 · Received May 22, 2025

Report

Report Number
1818910-2025-08327
Event Type
Malfunction
Date Received
May 22, 2025
Date of Event
April 29, 2025
Manufacturer
DEPUY IRELAND - 3015516266
Product Code
HSD
UDI-DI
10603295546610
PMA / PMN Number
K212737
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY ==> NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A MANUFACTURING RECORD EVALUATION (NC SEARCH) WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 550310040, LOT - 223758 AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED. AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT ==> A MANUFACTURING RECORD EVALUATION (NC SEARCH) WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 550310040, LOT - 223758 AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED

Description of Event or Problem · 0

IT WAS REPORTED THAT THE HEAD OF THE SCREW BROKE OFF UPON INSERTION DURING A PRIMARY SHOULDER REPAIR. THE BROKEN SCREW WAS LEFT IN PLACE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2321213 LOCKING SCREW 40 SHOULDER IMPLANT - LOCKING SCREW HSD DEPUY IRELAND - 3015516266 223758 10603295546610

Patients

Seq Age Sex Outcome Treatment
1 68 YR Male