MINIMED SURE-T
Report
- Report Number
- 3003442380-2025-09418
- Event Type
- Malfunction
- Date Received
- May 22, 2025
- Date of Event
- May 1, 2025
- Report Date
- June 25, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244019379
- PMA / PMN Number
- K160648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION, THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED). THE BATCH (B)(4) IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. COMPLAINT INVESTIGATIONS. THE REFERENCE SAMPLES FOR BATCH (B)(4) WERE PREVIOUSLY TESTED IN COMPLAINT (B)(4) ON 21 JUNE 2025. TEST RESULTS: VISUAL TEST ACCORDING TO WORK INSTRUCTION (WI) VERSION 3.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO WI VERSION 2.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: PACKING. THE LOT 6011047 WAS MANUFACTURED ACCORDING TO THE WI VERSION 99 AND PACKAGING IN THE MULTIVAC 14, ON 06 JANUARY 2025, WITH A TOTAL OF (B)(4) UNITS. GLUING OF TUBING. THE LOT 4M03028 WAS MANUFACTURED ACCORDING TO THE WI VERSION 65 IN THE GLUING OF TUBING IN THE MACHINE SC05, ON 15 DECEMBER 2024, WITH A TOTAL OF (B)(4) UNITS. THE LOT 4M03279 WAS MANUFACTURED ACCORDING TO THE WI VERSION 66 IN THE GLUING OF TUBING IN THE MACHINE SC05 & SC06, ON 03 JANUARY 2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 19 JUNE 2025 AGAINST MALFUNCTION CODE OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED) AND LOT 6011047 AND ANOTHER 1 COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO HARM, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ANOTHER 1 COMPLAINT RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE. NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON 01-MAY-2025. THE BLOCKAGE WAS IN THE TUBING. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1070358 | MINIMED SURE-T | UNO CONTACT DETACH G29 60/8 SC1 MIMX | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | MMT-874A | 6011047 | 05705244019379 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |