BD Q-SYTE CLOSED LUER ACCESS DEVICE
Report
- Report Number
- 1710034-2025-00874
- Event Type
- Malfunction
- Date Received
- May 21, 2025
- Date of Event
- March 22, 2025
- Report Date
- May 27, 2025
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FPA
- UDI-DI
- 00382903851003
- PMA / PMN Number
- K013621
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 385100 AND LOT NUMBER 4152203. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.
H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
NO ADDITIONAL INFORMATION.
IT WAS REPORTED THAT BD Q-SYTE CLOSED LUER ACCESS DEVICE SEPTUM DAMAGE WITH LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE TEAM LEADER REPLACED THE PICC SEPTUM WITH A NEEDLELESS INJECTION CAP. WHEN FLUSHING THE TUBING WITH SALINE SOLUTION, FLUID LEAKED FROM THE WHITE CONNECTOR. AFTER ADJUSTING THE TECHNIQUE AND RECONNECTING THE TUBING TIGHTLY, LEAKAGE PERSISTED, AND THE WHITE RUBBER AT THE CONNECTOR DID NOT REBOUND. NO HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 938246 | BD Q-SYTE CLOSED LUER ACCESS DEVICE | SET, ADMINISTRATION, INTRAVASCULAR | FPA | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 4152203 | 00382903851003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |