BD BACTEC¿ MGIT¿ 960 SIRE KIT
Report
- Report Number
- 1119779-2025-00460
- Event Type
- Malfunction
- Date Received
- May 21, 2025
- Date of Event
- April 23, 2025
- Report Date
- June 17, 2025
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- MJA
- UDI-DI
- 00382902451235
- PMA / PMN Number
- K003062
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
E1. INITIAL REPORTER PHONE #: (B)(6). THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA/510(K)#: K014123. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY - MGIT 960 SIRE SUPPLEMENT KIT BATCH 4242482 IS COMPOSED OF MGIT 960 SIRE SUPPLEMENT BATCH 4193147, MGIT 960 STREPTOMYCIN BATCH 4159881, MGIT 960 ISONIAZID BATCH 4159882, MGIT 960 RIFAMPIN BATCH 4159883, AND MGIT 960 ETHAMBUTOL BATCH 4159884. THE BATCH HISTORY RECORD REVIEWS FOR THE KIT AND EACH OF ITS COMPONENTS WERE SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON THIS KIT BATCH OR ANY OF THE COMPONENTS. MGIT 960 SIRE SUPPLEMENT IS MANUFACTURED BY REHYDRATING THE MEDIA COMPONENTS WITH USP PURIFIED WATER AND MIXED UNTIL A HOMOGENEOUS SOLUTION IS OBTAINED. THE SOLUTION IS THEN STERILE FILTERED AND DISPENSED INTO VIALS; STOPPERS ARE MANUALLY PLACED IN THE VIAL OPENING; SEPTUM CAPS ARE MANUALLY PLACED ON TOP OF THE STOPPER AND THEN MECHANICALLY CRIMPED PER STANDARD OPERATING PROCEDURES (SOP). ANTIBIOTICS MGIT 960 STREPTOMYCIN, MGIT 960 ISONIAZID, MGIT 960 RIFAMPIN AND MGIT 960 ETHAMBUTOL ARE MANUFACTURED BY REHYDRATING COMPONENTS IN USP PURIFIED WATER AND MIXING INTO A HOMOGENOUS SOLUTION. THE SOLUTION IS THEN STERILE FILTERED, DISPENSED INTO VIALS AND STOPPERED BY MACHINE. THE VIALS ARE LYOPHILIZED AND CRIMP CAPS ARE APPLIED PER SOP. EIGHT MGIT 960 SIRE SUPPLEMENT VIALS ARE THEN MANUALLY PACKAGED WITH ONE VIAL OF EACH ANTIBIOTIC TO MAKE A MGIT 960 SIRE SUPPLEMENT KIT (MATERIAL 245123). RETENTION SAMPLES FROM THE COMPONENTS FROM KIT BATCH 4242482 WERE AVAILABLE FOR INSPECTION. FOR THIS INVESTIGATION 12 RETENTION SAMPLES OF MGIT 960 SIRE SUPPLEMENT BATCH 4193147 WAS INSPECTED. NONE OF THE RETENTION SAMPLES CONTAINED CONTAMINATION OR FOREIGN MATTER. FOR FURTHER INVESTIGATION FOUR SIRE SUPPLEMENT VIALS WERE INCUBATED. TWO SUPPLEMENT VIAL CRIMP CAPS WERE REMOVED. ONE UNCRIMPED AND ONE CRIMPED SEALED SUPPLEMENT VIAL WERE PLACED INTO THE 20-25-DEGREE CELSIUS INCUBATOR. ONE UNCRIMPED AND ONE CRIMP SEALED SUPPLEMENT VIAL WERE PLACED INTO THE 33-37-DEGREES CELSIUS INCUBATOR. AT FOURTEEN DAYS OF INCUBATION NO MICROBIAL GROWTH WAS OBSERVED IN FOUR INCUBATED RETENTION VIALS. FOUR PHOTOS WERE RECEIVED TO ASSIST WITH THIS INVESTIGATION. ONE PHOTO SHOWED KIT CARTON LABEL BATCH 4242482, EXP 2025-12-03. ONE PHOTO SHOWED A STOPPER WITH SLIMY, BLACK MATERIAL ON THE UNDERSIDE OF THE STOPPER. ONE PHOTO SHOWED STOPPERED, UNCRIMPED SUPPLEMENT BATCH 4193147, EXP 2026-01-07. NO RETURNS WERE RECEIVED TO ASSIST WITH THIS INVESTIGATION. THIS COMPLAINT CAN BE CONFIRMED FOR CONTAMINATION. NO COMPLAINT TRENDS HAVE BEEN IDENTIFIED FOR THIS COMPLAINT. NO ACTIONS ARE INDICATED AT THIS TIME. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION.
IT WAS REPORTED BEFORE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, AN UNSPECIFIED NUMBER OF SIRE SUPPLEMENT VIALS CONTAINED BIOLOGICAL CONTAMINATION DESCRIBED AS "BLACK MOLD." THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
IT WAS REPORTED BEFORE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, AN UNSPECIFIED NUMBER OF SIRE SUPPLEMENT VIALS CONTAINED BIOLOGICAL CONTAMINATION DESCRIBED AS "BLACK MOLD." THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 724298 | BD BACTEC¿ MGIT¿ 960 SIRE KIT | SUSCEPTIBILITY TEST POWDERS, ANTIMYCOBACTERIAL | MJA | BECTON DICKINSON & CO. (SPARKS) | 4242482 | 00382902451235 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |