FDA Adverse Event Malfunction Summary report: N

INVOS

MDR report key: 22061230 · Received May 21, 2025

Report

Report Number
2936999-2025-00292
Event Type
Malfunction
Date Received
May 21, 2025
Date of Event
November 1, 2023
Report Date
May 21, 2025
Manufacturer
JIANGYIN SINBON ELECTRONICS CO LTD
Product Code
MUD
UDI-DI
10884521780125
PMA / PMN Number
K182868
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED AS PART OF REMEDIATION ACTIVITIES ASSOCIATED WITH CAPA # 643143, WHICH IS ADDRESSING MDR REPORTING GUIDANCE FROM FDAS 1995 PREAMBLE, WHICH ENSURES COMPLAINTS RELATED TO CORRECTIONS AND RECALLS WHICH WERE PREVIOUSLY REPORTED TO FDA UNDER 21 CFR 806 ARE NOW REPORTED AS MDRS; THIS IS NOT A NEW MALFUNCTION/EVENT. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ACCORDING TO THE REPORTER, DURING USE, THE DEVICE HAD TO BE SWITCHED TWICE BECAUSE RSO2 WAS PICKING UP THEN NOT SIGNAL ACQUISITION WAS INTERMITTENT. ANOTHER CABLE WAS USED AND SWAPPED TO ISOLATE FAILURE, AND THE PROBLEM WAS RESOLVED. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77713 INVOS OXIMETER, TISSUE SATURATION MUD JIANGYIN SINBON ELECTRONICS CO LTD PMAC71RSC 2023-04-28 10884521780125

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown