FDA Adverse Event Injury Summary report: N

DAVINCI 5

MDR report key: 22060366 · Received May 21, 2025

Report

Report Number
2955842-2025-21026
Event Type
Injury
Date Received
May 21, 2025
Date of Event
April 28, 2025
Report Date
April 28, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874119747
PMA / PMN Number
K232610
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT WAS ADDRESSED WITH PHONE SUPPORT. THE FIELD SERVICE ENGINEER (FSE) FOLLOWED WITH THE CLINICAL SALES REPRESENTATIVE (CSR). THE CSR COMMUNICATED ONCE THE DIGITAL VIDEO INTERFACE (DVI) CABLE WAS REPLACED THE IMAGE WORKED. NO FURTHER TROUBLE SHOOTING NEEDED. NO SITE VISIT WAS CONDUCTED. THE SYSTEM WAS WORKING PROPERLY, AND NO ADDITIONAL ACTION WAS REQUIRED AS THE ISSUE WAS RESOLVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED LIVER BIOPSY SURGICAL PROCEDURE, THE CLINICAL SALES REPRESENTATIVE (CSR) CONTACTED THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) TO REPORT THAT THE CUSTOMER REQUESTED ASSISTANCE GETTING A BK IMAGE TO POPULATE IN TILEPRO. THE CALLER HAS A HDMI/ VIDEO-IN, CONNECTION TO THE SURGEON SIDE CONSOLE 2 (SSC) TO VIEW THIRD-PARTY BK IMAGE. THE CUSTOMER WAS NOT ABLE TO GET THE BK TO RECOGNIZE. THE TSE CONFIRMED ALL TILEPRO SETTING AND CABLING WITH NO CHANGE TO THE TILEPRO VIEW. THE TSE THEN HAD THE CUSTOMER CONNECT THE BK DEVICE, POWER DOWN AND THEN BACK ON TO GET THE DA-VINCI TO RECOGNIZE TO NO AVAIL. THE TSE RECOMMENDED TRYING A NEW HDMI AND NEW BK DEVICE AS THE BK WAS NOT COMMUNICATING. THE OPERATING ROOM (OR) STAFF OPTED TO CONVERT TO OPEN WHILE THE ISSUES WAS BEING TROUBLESHOT. THE CSR REQUESTED A NEW HDMI AND POSSIBLE BK FROM THE CUSTOMER AND ENDED THE CALL. THE CUSTOMER MIGHT CALL BACK ONCE A REPLACEMENT BK AND HDMI CABLE WAS LOCATED. THE PROCEDURE WAS CONVERTED TO OPEN SURGERY. THE CSR CALLED BACK TO SEE IF THE VIDEO CONVERSION BOX, SHE BOUGHT AT A BIG BOX STORE WOULD WORK. THE CSR STATED THEY WERE TRYING TO HOOK UP VIDEO FROM THE BK ACTIVE WERE GOING TO TRY HOOK UP A DVI CABLE GOING FROM THE BK VIDEO OUTPUT THEN INTO THE CONVERSION BOX, THEN OUT OF THE CONVERSION BOX WITH A HDMI CABLE AND INTO THE CONSOLE HDMI INPUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
886497 DAVINCI 5 SURGEON SIDE CART NAY INTUITIVE SURGICAL, INC 380730-45 N/A 00886874119747

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention DA VINCI INSTRUMENTS AND ACCESSORIES