FDA Adverse Event Malfunction Summary report: N

MINIMED MIO ADVANCE

MDR report key: 22050542 · Received May 20, 2025

Report

Report Number
8021545-2025-01095
Event Type
Malfunction
Date Received
May 20, 2025
Date of Event
April 24, 2025
Report Date
November 14, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018877
PMA / PMN Number
K173879
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: FRANCE.

Additional Manufacturer Narrative · 0

THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER (8021545-2025-01095), WAS SUBMITTED ON 20-MAY-2025. UPON COMPLETION OF THE INVESTIGATION, THE MANUFACTURING DATE WAS UPDATED AS 05-NOV-2024. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS, H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010139. IN QUESTION WAS MANUFACTURED AT THE OSTED SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE 6010139 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) APPENDIX 1 BATCHCARD FOR PRODUCTION OF PACKAGING ROOM ON 05-NOV-2024 WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO PRODUCT WAS RETURNED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN FRANCE. IT WAS REPORTED THAT THE PATIENT EXPERIENCED THAT INFUSION SET WAS LEAKING FROM THE CONNECTION OF HUB TO THE CATHETER/RESERVOIR ON (B)(6) 2025. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2143387 MINIMED MIO ADVANCE UNO MIO ADV. GREY 60/6 SC1 10-PK INT FPA UNOMEDICAL A/S MMT-242A 6010139 05705244018877

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown