FDA Adverse Event Malfunction Summary report: N

YPSOPUMP INSET

MDR report key: 22050522 · Received May 20, 2025

Report

Report Number
3003442380-2025-09212
Event Type
Malfunction
Date Received
May 20, 2025
Date of Event
April 13, 2025
Report Date
May 6, 2025
Manufacturer
UNOMEDICAL UM-D
Product Code
FPA
PMA / PMN Number
K032854
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS H11: INVESTIGATION SUMMARY. THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6010690 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2 FOR THE OCCLUSION (E.G., OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED). (SOURCE/CAUSE CANNOT BE IDENTIFIED, ONLY USE WHEN A SPECIFIC MALFUNCTION CANNOT BE DETERMINED). IF THERE IS NO TROUBLESHOOTING, INCONCLUSIVE TROUBLESHOOTING, OR PARTIAL TROUBLESHOOT DONE AND IT CANNOT BE CONCLUDED TO BE INFUSION RELATED, REFER TO CANNOT BE DETERMINED CANNOT BE DETERMINED. NO ESCALATION REQUIRED. COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL AIR FLOW TEST 1 ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010690 WAS MANUFACTURED ACCORDING TO THE WI VERSION 75 MANUFACTURED IN THE LINE 5, ON 09/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED; NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 06/MAY/2025 AGAINST MALFUNCTION CODE OCCLUSION (E.G., OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED). (SOURCE/CAUSE CANNOT BE IDENTIFIED, ONLY USE WHEN A SPECIFIC MALFUNCTION CANNOT BE DETERMINED) AND LOT 6010690 AND NO MORE COMPLAINT HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR RETENTION SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION. NO FURTHER ACTIONS ARE REQUIRED. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN GERMANY. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INSULIN FLOW BLOCKED AND HIGH BLOOD GLUCOSE LEVELS ON (B)(6) 2025. IT WAS ALSO REPORTED THAT THE PATIENT WENT TO THE EMERGENCY ROOM (ER) AND WAS ADMITTED TO HOSPITAL FROM (B)(6) 2025 AND RECEIVED INTRAVENOUS (IV) FLUIDS WITH INSULIN SALINE AND THE PATIENT BLOOD GLUCOSE LEVELS WHILE ADMITTING THE HOSPITAL WAS 500 MG/DL. THE PATIENT EXPERIENCED KETONES WHICH WERE 3.8 MMOL/DL. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2142519 YPSOPUMP INSET UNO INSET II 60/6 GREY Y-CAP 10PAK INT FPA UNOMEDICAL UM-D 86-060-52B6 6010690

Patients

Seq Age Sex Outcome Treatment
1 NA Male Required Intervention