FDA Adverse Event Malfunction Summary report: N

ARCHITECT MAGNESIUM

MDR report key: 22047642 · Received May 19, 2025

Report

Report Number
3005094123-2025-00242
Event Type
Malfunction
Date Received
May 19, 2025
Date of Event
May 7, 2025
Report Date
May 23, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
JGJ
UDI-DI
00380740169862
PMA / PMN Number
K181748
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. E1 - POSTAL CODE: A LEADING ZERO WAS ADDED TO THE FIELD BECAUSE THE SYSTEM REQUIRES A MINIMUM OF 5 DIGITS IN THE POSTAL CODE FIELD.

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT MAGNESIUM, LIST NUMBER 03P68-34 , AND MANUFACTURING SITE: ABBOTT IRELAND DIAGNOSTICS DIVISION, LISNAMUCK, LONGFORD, IRELAND, N39E932, TO ARCHITECT C16000 PROCESSING MODULE, LIST NUMBER 03L77-01, ABBOTT LABORATORIES, 1915 HURD DRIVE, IRVING, TX, 75038. MDR NUMBER 3016438761-2025-00314 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED IMPRECISE MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE: 0.61-1.03 MMOL/L): INITIAL RESULT = 1.02 MMOL/L; REPEAT RESULT = 0.52 MMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED IMPRECISE MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE: 0.61-1.03 MMOL/L): INITIAL RESULT = 1.02 MMOL/L; REPEAT RESULT = 0.52 MMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2049772 ARCHITECT MAGNESIUM PHOTOMETRIC METHOD, MAGNESIUM JGJ ABBOTT IRELAND DIAGNOSTICS DIVISION 71411UD00 00380740169862

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ARC C16K PRC MOD, 03L77-01, (B)(6)| ARC C16K PRC MOD, 03L77-01, (B)(6)