ARCHITECT MAGNESIUM
Report
- Report Number
- 3005094123-2025-00242
- Event Type
- Malfunction
- Date Received
- May 19, 2025
- Date of Event
- May 7, 2025
- Report Date
- May 23, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- JGJ
- UDI-DI
- 00380740169862
- PMA / PMN Number
- K181748
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. E1 - POSTAL CODE: A LEADING ZERO WAS ADDED TO THE FIELD BECAUSE THE SYSTEM REQUIRES A MINIMUM OF 5 DIGITS IN THE POSTAL CODE FIELD.
AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT MAGNESIUM, LIST NUMBER 03P68-34 , AND MANUFACTURING SITE: ABBOTT IRELAND DIAGNOSTICS DIVISION, LISNAMUCK, LONGFORD, IRELAND, N39E932, TO ARCHITECT C16000 PROCESSING MODULE, LIST NUMBER 03L77-01, ABBOTT LABORATORIES, 1915 HURD DRIVE, IRVING, TX, 75038. MDR NUMBER 3016438761-2025-00314 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
THE CUSTOMER OBSERVED IMPRECISE MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE: 0.61-1.03 MMOL/L): INITIAL RESULT = 1.02 MMOL/L; REPEAT RESULT = 0.52 MMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
THE CUSTOMER OBSERVED IMPRECISE MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE: 0.61-1.03 MMOL/L): INITIAL RESULT = 1.02 MMOL/L; REPEAT RESULT = 0.52 MMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2049772 | ARCHITECT MAGNESIUM | PHOTOMETRIC METHOD, MAGNESIUM | JGJ | ABBOTT IRELAND DIAGNOSTICS DIVISION | 71411UD00 | 00380740169862 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ARC C16K PRC MOD, 03L77-01, (B)(6)| ARC C16K PRC MOD, 03L77-01, (B)(6) |