BD PHOENIX¿ PMIC/ID-107

MDR report key: 22047582 · Received May 19, 2025

Report

Report Number
1119779-2025-00393
Event Type
Malfunction
Date Received
May 19, 2025
Date of Event
April 25, 2025
Report Date
June 3, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
30382904486076
PMA / PMN Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B3: DATE OF EVENT IS UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. D4. MEDICAL DEVICE LOT#: UNKNOWN. D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K021954, K023273, K023301, K024152, K030677, K031306, K031679, K031679, K032131, K033784, K033907, K040006, K040106, K040716, K050089, K050555, K051689, K053241, K060214, K060217, K060218, K060493, K070809, K082538, K082852, AND K131331. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR MISIDENTIFICATION OF STAPHYLOCOCCUS AUREUS WHEN USING PHOENIX PANEL PMIC/ID-107 (CATALOG NUMBER 448607) BATCH NUMBER UNKNOWN. THE CUSTOMER DID NOT RETURN PANELS, ISOLATES OR PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE BATCH NUMBER WAS NOT PROVIDED; THEREFORE, THIS COMPLAINT IS UNCONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS COULD NOT BE PERFORMED SINCE A BATCH NUMBER WAS NOT PROVIDED. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ PMIC/ID-107 A PATIENT ISOLATE (STAPHYLOCOCCUS AUREUS) WAS MISIDENTIFIED AS GROUP B STREP OR STREPTOCOCCUS AGALACTIAE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 2 OF 2.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ PMIC/ID-107 A PATIENT ISOLATE (STAPHYLOCOCCUS AUREUS) WAS MISIDENTIFIED AS GROUP B STREP OR STREPTOCOCCUS AGALACTIAE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 2 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
158487 BD PHOENIX¿ PMIC/ID-107 SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION LON BECTON DICKINSON & CO. (SPARKS) UNKNOWN 30382904486076

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown