BD PHOENIX¿ PMIC/ID-107
Report
- Report Number
- 1119779-2025-00393
- Event Type
- Malfunction
- Date Received
- May 19, 2025
- Date of Event
- April 25, 2025
- Report Date
- June 3, 2025
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 30382904486076
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
B3: DATE OF EVENT IS UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. D4. MEDICAL DEVICE LOT#: UNKNOWN. D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K021954, K023273, K023301, K024152, K030677, K031306, K031679, K031679, K032131, K033784, K033907, K040006, K040106, K040716, K050089, K050555, K051689, K053241, K060214, K060217, K060218, K060493, K070809, K082538, K082852, AND K131331. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR MISIDENTIFICATION OF STAPHYLOCOCCUS AUREUS WHEN USING PHOENIX PANEL PMIC/ID-107 (CATALOG NUMBER 448607) BATCH NUMBER UNKNOWN. THE CUSTOMER DID NOT RETURN PANELS, ISOLATES OR PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE BATCH NUMBER WAS NOT PROVIDED; THEREFORE, THIS COMPLAINT IS UNCONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS COULD NOT BE PERFORMED SINCE A BATCH NUMBER WAS NOT PROVIDED. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ PMIC/ID-107 A PATIENT ISOLATE (STAPHYLOCOCCUS AUREUS) WAS MISIDENTIFIED AS GROUP B STREP OR STREPTOCOCCUS AGALACTIAE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 2 OF 2.
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ PMIC/ID-107 A PATIENT ISOLATE (STAPHYLOCOCCUS AUREUS) WAS MISIDENTIFIED AS GROUP B STREP OR STREPTOCOCCUS AGALACTIAE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 2 OF 2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 158487 | BD PHOENIX¿ PMIC/ID-107 | SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION | LON | BECTON DICKINSON & CO. (SPARKS) | UNKNOWN | 30382904486076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |