OMNIPOD DASH INSULIN MANAGEMENT SYSTEM
Report
- Report Number
- 3004464228-2025-21599
- Event Type
- Malfunction
- Date Received
- May 19, 2025
- Date of Event
- May 12, 2025
- Report Date
- May 19, 2025
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- UDI-DI
- 20385082000020
- PMA / PMN Number
- K211575
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. WE ARE UNABLE TO DETERMINE IF ANY PRODUCT CONDITION COULD HAVE CONTRIBUTED TO THE REPORTED HOSPITALIZATION AND HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED, AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS.
IT WAS REPORTED THE PATIENT'S BLOOD GLUCOSE LEVEL (BG) ROSE OVER 250 MG/DL WHILE WEARING THE POD BETWEEN 5 AND 24 HOURS. THE POD REPORTEDLY FELL OFF THE INFUSION SITE (LEG) WHILE ON THE PLAYGROUND, CAUSING THE CANNULA TO DISLODGE. AS TREATMENT, A NEW POD WAS PLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 132575 | OMNIPOD DASH INSULIN MANAGEMENT SYSTEM | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 18320 | PD1K09082431 | 20385082000020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR | Male |