FDA Adverse Event Malfunction Summary report: N

AUTOSOFT XC

MDR report key: 22041930 · Received May 19, 2025

Report

Report Number
3003442380-2025-09040
Event Type
Malfunction
Date Received
May 19, 2025
Date of Event
April 10, 2025
Report Date
July 29, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244016620
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6003535 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2 FOR THE CODE OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED) COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3 ON REFERENCE SAMPLES, REFERENCE SAMPLES PASSED THE TEST. FUNCTIONAL (FLOW) TEST ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, REFERENCE SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003535 WAS MANUFACTURED ACCORDING TO THE WI VERSION 108 MANUFACTURED IN THE LINE INSET 7, ON 04/OCT/2023, WITH A TOTAL OF (B)(4) UNITS. GLUING TUBING DHR REVIEW: THE LOT 3J03686 WAS MANUFACTURED ACCORDING TO THE WI VERSION 55 MANUFACTURED IN THE MACHINE SC01, ON 01/OCT/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 28/JUL/2025 AGAINST MALFUNCTION CODE OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED) AND LOT 6003535 AND NO OTHER COMPLAINT HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES RELATED TO THE COMPLAINT, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURED IN THE UNITED STATES. IT WAS REPORTED THAT, THE PATIENT ENCOUNTERED INFUSION SET'S BLOCK ALARM EVENT ON (B)(6) 2025. BLOCKAGE WAS AT TUBING. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2190528 AUTOSOFT XC UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1001680 6003535 05705244016620

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female