FDA Adverse Event Malfunction Summary report: N

TRUSTEEL

MDR report key: 22041757 · Received May 19, 2025

Report

Report Number
3003442380-2025-09066
Event Type
Malfunction
Date Received
May 19, 2025
Date of Event
April 25, 2025
Report Date
September 4, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244018426
PMA / PMN Number
K041545
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: REVISION 21 OF (B)(4) DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF (B)(4). COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011803, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011803 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 100 AND PACKAGED THE MULTIVAC M14 ON 25/FEB/2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING: LOT 5B00077 WAS MANUFACTURED ACCORDING TO THE WI VERSION 66 AND MANUFACTURED IN THE MACHINE SC08, ON 20/FEB/2025, WITH A TOTAL OF (B)(4) UNITS. LOT 5A04167 WAS MANUFACTURED ACCORDING TO THE WI VERSION 66 AND MANUFACTURED IN THE MACHINE SC08, ON 14/FEB/2025, WITH A TOTAL OF (B)(4) UNITS. LOT 5B03070 WAS MANUFACTURED ACCORDING TO THE WI VERSION 66 AND MANUFACTURED IN THE MACHINE SC05-SC06, ON 17/FEB/2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS IN THE TUBING. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2142572 TRUSTEEL UNO CONTACT DETACH G29 60/6TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1002833 6011803 05705244018426

Patients

Seq Age Sex Outcome Treatment
1 67 YR Female