TRUSTEEL
Report
- Report Number
- 3003442380-2025-09091
- Event Type
- Malfunction
- Date Received
- May 19, 2025
- Date of Event
- April 19, 2025
- Report Date
- October 14, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018488
- PMA / PMN Number
- K041545
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: REVISION 21 OF (B)(4) DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF (B)(4). COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010326, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010326 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 98 AND PACKAGING IN THE MACHINE MULTIVAC 14 ON 24-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING OF THE LOT 4L03190 WAS MANUFACTURED ACCORDING TO THE WI VERSION 34 AND MANUFACTURED IN THE MACHINE LS24-LS25, ON 15-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING OF THE LOT 4K02964 WAS MANUFACTURED ACCORDING TO THE WI VERSION 34 AND MANUFACTURED IN THE MACHINE LS06-LS07, ON 14-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING OF THE LOT 4K05788 WAS MANUFACTURED ACCORDING TO THE WI VERSION 34 AND MANUFACTURED IN THE MACHINE LS06-LS07, ON 24-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING OF THE LOT 4L04889 WAS MANUFACTURED ACCORDING TO THE WI VERSION 34 AND MANUFACTURED IN THE MACHINE LS24-LS25, ON 24-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING CONNECTOR OF THE LOT 4K05776 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3, ON 14-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING CONNECTOR OF THE LOT 4K05519 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3, ON 14-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING CONNECTOR OF THE LOT 4L03430 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3, ON 25-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING CONNECTOR OF THE LOT 4L03431 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3, ON 26-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING CONNECTOR OF THE LOT 4K05777 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3, ON 24-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN EVENT OF INSULIN FLOW BLOCKED ON (B)(6) 2025. TROUBLESHOOTING CONFIRMED BLOCKAGE AT THE SITE. NO FURTHER INFORMATION WAS AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2050152 | TRUSTEEL | UNO CONTACT DETACH G29 60/8TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1002835 | UNKNOWN | 05705244018488 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Female |