FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2204020 · Received August 11, 2011

Report

Report Number
2649622-2011-12629
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S17
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. HIGH RESISTANCE/IMPEDANCE WAS NOTED AS TWO PATIENT ALERTS FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE OCCURRED ON (B)(6) 2011 03:00:13 AND (B)(6) 2011 03:00:11. THE WEEKLY PACE IMPEDANCE TREND DATA SHOWS A GRADUAL INCREASE FOR MIN AND MAX VENTRICULAR PACE BI IMPEDANCE FROM 627 TO 3192 OHMS PEAK BETWEEN (B)(6) 2010 AND (B)(6) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS APPARENTLY FRACTURED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6944 ASKU

Patients

Seq Age Sex Outcome Treatment
1 57 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD