CI-1600-05
Report
- Report Number
- 3006556115-2025-00626
- Event Type
- Malfunction
- Date Received
- May 16, 2025
- Date of Event
- June 24, 2025
- Report Date
- April 25, 2025
- Manufacturer
- ADVANCED BIONICS, LLC
- Product Code
- MCM
- UDI-DI
- 07630016855696
- PMA / PMN Number
- P960058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CI
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE. DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
ADDITIONAL INFORMATION SECTION H.6, D.9. ADVANCED BIONICS CONSIDERS THE INVESTIGATION INTO THIS REPORTABLE EVENT AS CLOSED. THE EXTERNAL VISUAL INSPECTION REVEALED SLICED SILICONE OVERMOLD ON THE TOP AND BOTTOM COVERS, AS WELL AS A SEVERED ELECTRODE. THIS IS BELIEVED TO HAVE OCCURRED DURING REVISION SURGERY. THE DEVICE PASSED PHOTOGRAPHIC IMAGING INSPECTION. SYSTEM LOCK WAS VERIFIED. THE CONDITION OF THE ELECTRODE PREVENTED AN ELECTRICAL TEST FROM BEING PERFORMED. THE DEVICE PASSED SOME OF THE ELECTRICAL TESTS PERFORMED. THE FAILURE OF THIS DEVICE IS ATTRIBUTED TO AN ELECTRODE SHORT IN THE ELECTRODE POCKET. A CORRECTIVE ACTION WAS IMPLEMENTED. THIS VERSION OF THE HIRES ULTRA DEVICE IS NO LONGER DISTRIBUTED. THIS IS THE FINAL REPORT. DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
THE RECIPIENT IS REPORTEDLY EXPERIENCING DECREASED PERFORMANCE. A REVIEW OF THE RECIPIENT'S TEST DATA INDICATES IMPEDANCE ISSUES. REVISION SURGERY IS SCHEDULED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2129333 | CI-1600-05 | COCHLEAR IMPLANT | MCM | ADVANCED BIONICS, LLC | HIRES¿ ULTRA IMPLANT | NA | 07630016855696 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Female |