FDA Adverse Event Malfunction Summary report: N

CI-1600-05

MDR report key: 22040077 · Received May 16, 2025

Report

Report Number
3006556115-2025-00626
Event Type
Malfunction
Date Received
May 16, 2025
Date of Event
June 24, 2025
Report Date
April 25, 2025
Manufacturer
ADVANCED BIONICS, LLC
Product Code
MCM
UDI-DI
07630016855696
PMA / PMN Number
P960058
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.

Additional Manufacturer Narrative · 0

THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE. DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION SECTION H.6, D.9. ADVANCED BIONICS CONSIDERS THE INVESTIGATION INTO THIS REPORTABLE EVENT AS CLOSED. THE EXTERNAL VISUAL INSPECTION REVEALED SLICED SILICONE OVERMOLD ON THE TOP AND BOTTOM COVERS, AS WELL AS A SEVERED ELECTRODE. THIS IS BELIEVED TO HAVE OCCURRED DURING REVISION SURGERY. THE DEVICE PASSED PHOTOGRAPHIC IMAGING INSPECTION. SYSTEM LOCK WAS VERIFIED. THE CONDITION OF THE ELECTRODE PREVENTED AN ELECTRICAL TEST FROM BEING PERFORMED. THE DEVICE PASSED SOME OF THE ELECTRICAL TESTS PERFORMED. THE FAILURE OF THIS DEVICE IS ATTRIBUTED TO AN ELECTRODE SHORT IN THE ELECTRODE POCKET. A CORRECTIVE ACTION WAS IMPLEMENTED. THIS VERSION OF THE HIRES ULTRA DEVICE IS NO LONGER DISTRIBUTED. THIS IS THE FINAL REPORT. DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.

Description of Event or Problem · 0

THE RECIPIENT IS REPORTEDLY EXPERIENCING DECREASED PERFORMANCE. A REVIEW OF THE RECIPIENT'S TEST DATA INDICATES IMPEDANCE ISSUES. REVISION SURGERY IS SCHEDULED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2129333 CI-1600-05 COCHLEAR IMPLANT MCM ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT NA 07630016855696

Patients

Seq Age Sex Outcome Treatment
1 55 YR Female