FDA Adverse Event Malfunction Summary report: N

CENTERED HEAD CRCO 48X18

MDR report key: 22039760 · Received May 16, 2025

Report

Report Number
3014128390-2025-00024
Event Type
Malfunction
Date Received
May 16, 2025
Date of Event
May 6, 2025
Report Date
May 16, 2025
Manufacturer
FX SHOULDER SOLUTIONS DALLAS
Product Code
HSD
UDI-DI
03701037300428
PMA / PMN Number
K123814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

WALL OF PLASTIC TRAY WAS DAMAGED, COMPROMISING THE SEAL. CAPA-2024-0039 HAS BEEN OPENED TO ADDRESS THIS ISSUE.

Description of Event or Problem · 0

IN A CASE WITH THE SURGEON ON (B)(6) 2025, A 48X18 CENTERED HEAD WAS OPENED DURING SURGERY AND THE TRAY WAS SEEN TO BE CRACKED IN A MANNER THAT BREACHES THE STERILE BARRIER. THE TRAY HAS CRACKED IN THE CORNER OPPOSITE PEELING CORNER, RESULTING IN CHIPPING OF THE SEALING LIP. THE TYVEK BARRIER WAS PEELED BACK APPROXIMATELY 1 INCH AS A RESULT OF THE DAMAGE. THE COMPONENT ITSELF, AND THE PLASTIC WRAPPED SEAL, APPEAR TO NOT HAVE BEEN COMPROMISED. NO SURGICAL DELAY WAS NOTED. CAPA-2024-0039 HAS BEEN OPENED TO ADDRESS THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2050229 CENTERED HEAD CRCO 48X18 SHOULDER PROSTHESIS HSD FX SHOULDER SOLUTIONS DALLAS P1660 03701037300428

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown