FDA Adverse Event Injury Summary report: N

PERCLOSE¿ PROSTYLE¿

MDR report key: 22039602 · Received May 16, 2025

Report

Report Number
2024168-2025-05164
Event Type
Injury
Date Received
May 16, 2025
Date of Event
April 18, 2025
Report Date
July 2, 2025
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648344039
PMA / PMN Number
P960043
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL DEVICES REFERENCED IN B5 ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.

Additional Manufacturer Narrative · 0

A VISUAL INSPECTION, FUNCTIONAL TESTING, DIMENSIONAL ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE FAILURE TO CYCLE WAS CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. OPERATIONAL CONTEXT/USER ERROR/ PQI: BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED ISSUE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. IT IS LIKELY A POSTERIOR NEEDLE TO CUFF MISS OCCURRED. THE POSTERIOR NEEDLE LIKELY STRUCK THE FOOT CAUSING SEPARATION OF THE FOOT. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY USING THREE PROSTYLE DEVICES AFTER A PERIPHERAL ANGIOPLASTY PROCEDURE USING AN 8F SHEATH. REPORTEDLY, IT WAS NOTED THAT THE NEEDLES DID NOT FIX CORRECTLY AND THE TWO WIRES [SUTURES] DID NOT COME OUT IN STEP 3. A SECOND AND THIRD PROSTYLE DEVICE WERE ATTEMPTED BUT THE SAME THING HAPPENED. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2190655 PERCLOSE¿ PROSTYLE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12773-02 4061841 08717648344039

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention