FDA Adverse Event Injury Summary report: N

KAPPA 700 DR

MDR report key: 2203842 · Received August 11, 2011

Report

Report Number
2647346-2011-01051
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035
Removal / Correction Number
Z-1510/1519-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION AND THE DEVICE MET EXPECTED LONGEVITY.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD HAD ELEVATED IMPEDANCE CONSISTENT WITH A LEAD FRACTURE AND THERE WAS A LEAD MALFUNCTION. THE LEAD WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED BY THE PATIENT THAT THE "[BATTERY] LIFE DROPPED 60%" DURING A 5 MONTH PERIOD. IT WAS LATER CONFIRMED BY THE PHYSICIAN THAT THE DEVICE INDICATED ELECTIVE REPLACEMENT NORMALLY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KAPPA 700 DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. KDR703 ASKU

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD