FDA Adverse Event
Injury
Summary report: N
SENSIA DR
MDR report key: 2203826
·
Received August 11, 2011
Report
- Report Number
- 6000144-2011-04351
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NVZ
- PMA / PMN Number
- P980035/S043
- Removal / Correction Number
- Z-2120-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS DEVICE WAS INCLUDED IN A FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE INDICATED NORMAL BATTERY DEPLETION.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION AND THE DEVICE MET EXPECTED LONGEVITY.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE HAD PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SENSIA DR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | SEDR01 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Hospitalization| R | 4024 IMPLANTABLE PACING LEAD| 4524 IMPLANTABLE PACING LEAD |