CAPSUREFIX NOVUS MRI
Report
- Report Number
- 2649622-2011-12604
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- June 8, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED AND PRIMARY ANALYSIS REVEALED NO ANOMALIES. THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED). THE OUTER INSULATION WAS BREACHED CUT AND THERE WAS BLOOD PRESENT IN/ON THE HELIX/LOBE MECHANISM. THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.
IT WAS REPORTED THAT DURING IMPLANT, THE ATRIAL LEAD HAD MULTIPLE DISLODGEMENTS. THE LEAD WAS NOT USED, AND A NEW LEAD IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Other |