FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2203684 · Received August 11, 2011

Report

Report Number
2649622-2011-11888
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 24, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, THE PHYSICIAN HAD TROUBLE GETTING THE RIGHT ATRIAL LEAD TO MAKE A "J" SHAPE IN THE ATRIUM WITH THE STYLET. THE LEAD WAS REMOVED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 87 YR Other (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) LEAD| (B)(4) NEURO EXTENSION