FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2203673 · Received August 11, 2011

Report

Report Number
2649622-2011-12364
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
June 2, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYSIS REVEALED THAT THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS ALSO NOTED THAT THE INNER INSULATION WAS KINKED/BUCKLED, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND SLEEVE HEAD AND THE LEAD APPEARED DAMAGE AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE HELIX WOULD NOT EXTEND. THE LEAD WAS NOT USED. A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Other