FDA Adverse Event Injury Summary report: N

CAPSURE SP

MDR report key: 2203609 · Received August 11, 2011

Report

Report Number
2649622-2011-11886
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S12
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DISTAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED. THE INNER INSULATION WAS BREACHED DUE TO METAL ION OXIDATION (MIO), AND ALL INSULATORS EXHIBITED COSMETIC MIO. THE OUTER INSULATION EXHIBITED COSMETIC ENVIRONMENTAL STRESS CRACKING (ESC).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD HAD IMPEDANCE THAT WAS TRENDING DOWNWARD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4524 ASKU

Patients

Seq Age Sex Outcome Treatment
1 89 YR Hospitalization| R 4194 IMPLANTABLE PACING LEAD| 7299 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 7299 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD| 6949 IMPLANTABLE TACHY LEAD