CAPSURE SP
Report
- Report Number
- 2649622-2011-11886
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P830061/S12
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DISTAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED. THE INNER INSULATION WAS BREACHED DUE TO METAL ION OXIDATION (MIO), AND ALL INSULATORS EXHIBITED COSMETIC MIO. THE OUTER INSULATION EXHIBITED COSMETIC ENVIRONMENTAL STRESS CRACKING (ESC).
IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD HAD IMPEDANCE THAT WAS TRENDING DOWNWARD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SP | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4524 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 89 YR | Hospitalization| R | 4194 IMPLANTABLE PACING LEAD| 7299 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 7299 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD| 6949 IMPLANTABLE TACHY LEAD |