FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 2203591 · Received August 11, 2011

Report

Report Number
6000144-2011-04195
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE DEVICE WAS RETURNED AND ANALYZED. CALCULATIONS BASED ON IMPLANT PARAMETERS INDICATE THAT THE DEVICE HAD MET ITS 99.9% EXPECTED LONGEVITY. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. THIS DEVICE IS PART OF THE FIELD ACTION AND HAS TESTED CONSISTENT WITH THE FIELD ACTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS REPLACED DUE TO ERI (ELECTIVE REPLACEMENT INDICATOR). THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Hospitalization| R 5568 IMPLANTABLE PACING LEAD| 4076 IMPLANTABLE PACING LEAD