FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 2203574 · Received August 11, 2011

Report

Report Number
6000144-2011-04554
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD A POWER ON RESET (POR) THAT WAS ASSOCIATED WITH RECENT RADIATION THERAPY. THE DEVICE RESET BACK TO PREVIOUS PACING PARAMETERS WITHOUT INCIDENT. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 81 YR Required Intervention 5076 IMPLANTABLE PACING LEAD| 5568 IMPLANTABLE PACING LEAD