ENRHYTHM DR
Report
- Report Number
- 6000144-2011-04193
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): CALCULATIONS BASED ON IMPLANT PARAMETERS INDICATE THAT THE DEVICE HAD MET ITS 99.9% EXPECTED LONGEVITY. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. THIS DEVICE IS PART OF THE FIELD ACTION AND HAS TESTED CONSISTENT WITH THE FIELD ACTION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. NO IMPEDANCE ANOMALIES WERE FOUND. AS OF (B)(6) 2011, BATTERY IMPEDANCE WAS WITHIN NORMAL LIMITS. NO BATTERY VOLTAGE ANOMALIES WERE FOUND. AS OF (B)(6) 2011, BATTERY VOLTAGE APPEARED NORMAL AND STABLE AT 3.00 V. RAMWARE VERSION 8 INSTALLED ON (B)(6) 2011.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): CALCULATIONS BASED ON IMPLANT PARAMETERS INDICATE THAT THE DEVICE HAD MET ITS 99.9% EXPECTED LONGEVITY. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. THIS DEVICE IS PART OF THE FIELD ACTION AND HAS TESTED CONSISTENT WITH THE FIELD ACTION.
IT WAS REPORTED THAT THE DEVICE REACHED ERI (ELECTIVE REPLACEMENT INDICATOR) AND THAT EARLY BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | P1501DR | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Hospitalization| R | 5054 IMPLANTABLE PACING LEAD| 5054 IMPLANTABLE PACING LEAD| 5554 IMPLANTABLE PACING LEAD| 5554 IMPLANTABLE PACING LEAD |