FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2203572 · Received August 11, 2011

Report

Report Number
2649622-2011-11884
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT FELL A FEW WEEKS AGO AND STARTED EXPERIENCING DIAPHRAGMATIC STIMULATION WITH RIGHT VENTRICULAR PACING. THE LEAD TESTED WITHIN NORMAL LIMITS; HOWEVER, THE PHYSICIAN CHOSE TO EXPLANT THE LEAD AND PLACE A NEW LEAD IN A DIFFERENT POSITION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 73 YR Hospitalization| R ADDR01 IMPLANTABLE PULSE GENERATOR| 5568 IMPLANTABLE PACING LEAD