SOL-MILLENNIUM
Report
- Report Number
- 3014312726-2024-00192
- Event Type
- Malfunction
- Date Received
- May 16, 2025
- Date of Event
- January 26, 2024
- Report Date
- August 13, 2025
- Manufacturer
- SOL-MILLENNIUM MEDICAL INC.
- Product Code
- QNQ
- UDI-DI
- 00818392018165
- PMA / PMN Number
- K242291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE IS NO EVIDENCE OF QUALITY ISSUES ASSOCIATED WITH THIS COMPLAINT BASED ON THE ARCHIVE SAMPLE ANALYSIS. BASED ON THE DETERMINED RISK PRIORITY NUMBER, THE RESIDUAL RISK RELATED TO THE COMPLAINT IS CONSIDERED ACCEPTABLE. SOL MILLENNIUM CONTINUES TO MONITOR THIS KIND OF ISSUE. ADDITIONAL ASSESSMENT WILL BE TAKEN IF A SIGNIFICANT PATTERN EMERGES. UNCONFIRMED. THIS REPORT WAS INITIALLY SUBMITTED ON 04/08/2024. DUE TO ISSUES WITH DISPLAYING THE INITIAL SUBMISSION, THIS SUBMISSION CONTAINS BOTH INITIAL AND SUPPLEMENTAL INFORMATION.
THIS REPORT WHEN INITIALLY SUBMITTED ENCOUNTERED AN INTERNAL ISSUE POTENTIALLY RELATED TO DATABASE CONNECTIVITY. THIS ISSUE WAS IDENTIFIED WHILE IMPLEMENTING CAPA-57 WHICH WAS OPENED FOLLOWING AN FDA INSPECTION.
"CUSTOMER REPORTED: WE RECEIVED A COMPLAINT FROM RIVM FOR ITEM P180001PP WITH BATCH 04105025. IT HAD SEVERAL DEFECTS: SYRINGE WITHOUT GRADUATION MARKS: 1 PIECE. SYRINGE WITHOUT PLUNGER: 1 PIECE. CLOSED PACKAGING WITHOUT SYRINGE: 1 PIECE. SYRINGE EXPANDED AND WITH FISSURES; FRACTURE CRACKS: 3 PIECES."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 132442 | SOL-MILLENNIUM | SOL-M 1ML LUER LOCK SYRINGE W/O NEEDLE | QNQ | SOL-MILLENNIUM MEDICAL INC. | P180001PP | 04105025 | 00818392018165 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |