FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2203546 · Received August 11, 2011

Report

Report Number
2649622-2011-12352
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
June 1, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE FULL LEAD FOUND THE OUTER INSULATION KINKED/BUCKLED. FURTHERMORE, IT WAS NOTED THAT THE DISTAL CONDUCTOR CONNECTOR PIN WAS BENT, ALL CONDUCTORS WERE STRETCHED, THE LEAD WAS STRETCHED, AND THE LEAD APPEARED DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT OF THE LEAD, THE PHYSICIAN TRIED TO PASS THE LEAD INTO THE INTRODUCER, BUT EXPERIENCED DIFFICULTIES. IT WAS NOTED THAT THERE WAS INSULATION DAMAGE OF THE LEAD. THE LEAD WAS NOT USED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other