FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 2203541 · Received August 11, 2011

Report

Report Number
2649622-2011-12355
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE DISTAL SEGMENT WAS RETURNED AND ANALYZED AND THE INNER AND OUTER INSULATION HAD BREACHED METAL INDUCED OXIDATION. THERE WAS BLOOD/ BODY FLUID (NOT OBSTRUCTED) ON ALL CONDUCTORS AND THE CONDUCTORS WERE DISTORTED. THE OUTER INSULATION WAS MELTED AND HAD COSMETIC ENVIRONMENTAL STRESS CRACKING. THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THE TIP SEAL WAS OBSERVED OUT OF POSITION.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE EVALUATION SUMMARY FOR (B)(4): THE DISTAL SEGMENT WAS RETURNED AND ANALYZED AND THE INNER AND OUTER INSULATION HAD BREACHED METAL INDUCED OXIDATION. THERE WAS BLOOD/ BODY FLUID (NOT OBSTRUCTED) ON ALL CONDUCTORS AND THE CONDUCTORS WERE DISTORTED. THE OUTER INSULATION WAS MELTED AND HAD COSMETIC ENVIRONMENTAL STRESS CRACKING. THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THE TIP SEAL WAS OBSERVED OUT OF POSITION.

Description of Event or Problem · 1

IT WAS REPORTED THE ATRIAL LEAD HAD INCREASED THRESHOLDS AND HIGH IMPEDANCE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4568 ASKU

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| R 6949 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD| 4193 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4193 IMPLANTABLE PACING LEAD