FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 2203537 · Received August 11, 2011

Report

Report Number
2649622-2011-11879
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 18, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED AND ANALYSIS REVEALED ALL CONDUCTORS WERE DISTORTED. THE LEAD WAS DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT THE WIRE WAS NOT ABLE TO BE PASSED THROUGH THE LEFT VENTRICULAR LEAD. THE LEAD WAS NOT IMPLANTED AND A DIFFERENT LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 60 YR Other 5076 IMPLANTABLE PACING LEAD