SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-12350
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- April 7, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY:(B)(4) THE FULL LEAD WAS RETURNED AND ANALYSIS REVEALED THAT THE DISTAL CONDUCTOR HAD BLOOD/FLUID OBSTRUCTION. IT WAS ALSO NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.
IT WAS REPORTED DURING THE IMPLANT PROCEDURE IT WAS DIFFICULT "TO INTRODUCE THE WIRE INSIDE THE LUMEN". A NEW LEAD WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT WAS LISTED AS "DOING WELL". NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Other |