FDA Adverse Event
Malfunction
Summary report: N
CAPSURE Z NOVUS
MDR report key: 2203467
·
Received August 11, 2011
Report
- Report Number
- 2649622-2011-12118
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- June 9, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P850089/S38
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): THE DISTAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE LEAD PACING THRESHOLD WAS MORE THAN 8 V. THE LEAD WAS NOT USED AND A SECOND LEAD WAS ATTEMPTED. THE PACING THRESHOLD OF THE SECOND LEAD WAS MORE THAN 8 V ALSO, HOWEVER THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE Z NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5554 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |