FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2203461 · Received August 11, 2011

Report

Report Number
2649622-2011-11702
Event Type
Injury
Date Received
August 11, 2011
Date of Event
May 12, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND PRIMARY ANALYSIS REVEALED ANOMALIES FOUND. THE PROXIMAL CONDUCTOR WAS DISTORTED AND BLOOD WAS PRESENT IN/ON THE HELIX/LOBE MECHANISM. THERE WAS ALSO APPARENT EXPLANT DAMAGE. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND PRIMARY ANALYSIS REVEALED NO ANOMALIES. BLOOD WAS PRESENT IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS DISLODGMENT OF THE ATRIAL LEAD AND HIGH THRESHOLD ON THE VENTRICULAR LEAD. IT WAS ALSO REPORTED THAT THE PATIENT EXPERIENCED PHRENIC NERVE STIMULATION. THE LEADS WERE EXTRACTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI ASKU

Patients

Seq Age Sex Outcome Treatment
1 49 YR Hospitalization| R